Skip navigation

Zastosuj identyfikator do podlinkowania lub zacytowania tej pozycji: http://hdl.handle.net/20.500.12128/445
Pełny rekord metadanych
DC poleWartośćJęzyk
dc.contributor.authorLopes, Bernardo T.-
dc.contributor.authorRoberts, Cynthia-
dc.contributor.authorElsheikh, Ahmed-
dc.contributor.authorVinciguerra, Riccardo-
dc.contributor.authorVinciguerra, Paolo-
dc.contributor.authorReisdorf, Sven-
dc.contributor.authorBerger, Stefanie-
dc.contributor.authorKoprowski, Robert-
dc.contributor.authorAmbrosio, Renato (Jr.)-
dc.date.accessioned2017-11-25T19:13:46Z-
dc.date.available2017-11-25T19:13:46Z-
dc.date.issued2017-
dc.identifier.citationJournal of Ophthalmology, Art. no. 8515742, (2017), s. 1-4pl_PL
dc.identifier.issn2090-004X-
dc.identifier.issn2090-0058-
dc.identifier.urihttp://hdl.handle.net/20.500.12128/445-
dc.description.abstractPurpose. To assess the repeatability and reproducibility of dynamic corneal response parameters measured by the Corvis ST (Oculus, Wetzlar, Germany). Methods. One eye randomly selected from 32 healthy volunteers was examined by the Corvis ST. Three different devices were used in an alternated random order for taking three measurements at each device in each subject. Standard intraocular pressure (IOP), the biomechanical-compensated IOP (bIOP), and DCR parameters were evaluated. The within-subject standard deviation (zeta w) and coefficient of variation (CV) were assessed. Results. Regarding pressure indices, the zeta w was below 1 mmHg for repeatability (0.98 for IOP and 0.89 for bIOP) and the CV was 6.6% for IOP and 6.1% for bIOP. For reproducibility, the zeta w was around 1 mmHg (1.12 for IOP and 1.05 for bIOP) and the CV was 7.6% for IOP and 7.1% for bIOP. Most of DCR indices presented CV for repeatability below 4%. For reproducibility, the CV of most of the indices were below 6%. The deformation amplitude (DA) ratio in 1 mm and integrated radius were below 4% (1.2% and 3.8%, resp.). Conclusions. The Corvis ST showed good precision (repeatability and reproducibility) for IOP measurements and for DCR in healthy eyes.pl_PL
dc.language.isoenpl_PL
dc.rightsUznanie autorstwa 3.0 Polska*
dc.rights.urihttp://creativecommons.org/licenses/by/3.0/pl/*
dc.subjectForme-Fruste Keratoconuspl_PL
dc.subjectBiomechanical Propertiespl_PL
dc.subjectAnalyzerpl_PL
dc.subjectEctasiapl_PL
dc.subjectCorvis STpl_PL
dc.titleRepeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis STpl_PL
dc.typeinfo:eu-repo/semantics/articlepl_PL
dc.identifier.doi10.1155/2017/8515742-
Pojawia się w kolekcji:Artykuły (WNŚiT)

Pliki tej pozycji:
Plik Opis RozmiarFormat 
Lopes_Repeatability_and_Reproducibility_of_Intraocular_Pressure.pdf361,64 kBAdobe PDFPrzejrzyj / Otwórz
Pokaż prosty rekord


Uznanie Autorstwa 3.0 Polska Creative Commons Creative Commons