Tytuł: Experimental observation of an extremely high electron lifetime with the ICARUS-T600 LAr-TPC
Autor: Antonello, M.
Baibussinov, B.
Benetti, P.
Boffelli, F.
Bubak, Arkadiusz
Calligarich, E.
Centro, S.
Cesana, A.
Cieślik, K.
Cline, D. B.
Cocco, A. G.
Dabrowska, A.
Dequal, D.
Dermenev, A.
Dolfini, R.
Falcone, A.
Farnese, C.
Fava, A.
Ferrari, A.
Fiorillo, G.
Gibin, D.
Gninenko, S.
Guglielmi, A.
Haranczyk, M.
Holeczek, Jacek
Kirsanov, M.
Kisiel, Jan
Kochanek, Izabela
Lagoda, J.
Mania, Sławomir
Menegolli, A.
Meng, G.
Montanari, C.
Nicoletto, M.
Otwinowski, S.
Picchi, P.
Pietropaolo, F.
Płoński, P.
Rappoldi, A.
Raselli, G. L.
Rossella, M.
Rubbia, C.
Sala, P.
Scaramelli, A.
Segreto, E.
Sergiampietri, F.
Stefan, D.
Sulej, R.
Szarska, M.
Terrani, M.
Torti, M.
Varanini, F.
Ventura, S.
Vignoli, C.
Wang, H.
Yang, X.
Zalewska, A.
Zani, A.
Zaremba, K.
Słowa kluczowe: charge transport and multiplication in liquid media; neutrino detectors; cryogenic detectors; time projection chambers
Data wydania: 2014
Źródło: Journal of Instrumentation, Vol. 9, iss. 12 (2014), art. no. P12006
Abstrakt: The ICARUS T600 detector, the largest liquid Argon Time Projection Chamber (LAr-TPC) realized after many years of R&D activities, was installed and successfully operated for 3 years at the INFN Gran Sasso underground Laboratory. One of the most important issues was the need of an extremely low residual electronegative impurity content in the liquid Argon, in order to transport the free electrons created by ionizing particles with very small attenuation along the drift path. The solutions adopted for the Argon recirculation and purification systems have permitted to reach impressive results in terms of Argon purity and a free electron lifetime exceeding 15 ms, corresponding to about 20 parts per trillion of O 2 -equivalent contamination, a milestone for any future project involving LAr-TPCs and the development of higher detector mass scales.
URI: http://hdl.handle.net/20.500.12128/9108
DOI: 10.1088/1748-0221/9/12/P12006
ISSN: 1748-0221
Pojawia się w kolekcji:Artykuły (WNŚiT)

Pliki tej pozycji:
Plik Opis RozmiarFormat 
Antonello_Experimental_observation_of_an_extremely.pdf1,24 MBAdobe PDFPrzejrzyj / Otwórz
Pokaż pełny rekord


Uznanie Autorstwa 3.0 Polska Creative Commons Creative Commons