Skip navigation

Zastosuj identyfikator do podlinkowania lub zacytowania tej pozycji: http://hdl.handle.net/20.500.12128/12700
Tytuł: Bioresorbable Stent in Anterior Cruciate Ligament Reconstruction
Autor: Ficek, Krzysztof
Rajca, Jolanta
Stolarz, Mateusz
Stodolak-Zych, Ewa
Wieczorek, Jarosław
Muzalewska, Małgorzata
Wyleżoł, Marek
Wróbel, Zygmunt
Binkowski, Marcin
Błażewicz, Stanisław
Słowa kluczowe: anterior cruciate ligament reconstruction; bone tunnel enlargement; X-raymicrotomography; polylactide
Data wydania: 2019
Źródło: Polymers, Vol. 11 (2019), Art. No. 1961
Abstrakt: The exact causes of failure of anterior cruciate ligament (ACL) reconstruction are still unknown. A key to successful ACL reconstruction is the prevention of bone tunnel enlargement (BTE). In this study, a new strategy to improve the outcome of ACL reconstruction was analyzed using a bioresorbable polylactide (PLA) stent as a catalyst for the healing process. The study included 24 sheep with 12 months of age. The animals were randomized to the PLA group (n = 16) and control group (n = 8), subjected to the ACL reconstruction with and without the implantation of the PLA tube, respectively. The sheep were sacrificed 6 or 12 weeks post-procedure, and their knee joints were evaluated by X-ray microcomputed tomography with a 50 m resolution. While the analysis of tibial and femoral tunnel diameters and volumes demonstrated the presence of BTE in both groups, the enlargement was less evident in the PLA group. Also, the microstructural parameters of the bone adjacent to the tunnels tended to be better in the PLA group. This suggested that the implantation of a bioresorbable PLA tube might facilitate osteointegration of the tendon graft after the ACL reconstruction. The beneficial e ects of the stent were likely associated with osteogenic and osteoconductive properties of polylactide.
URI: http://hdl.handle.net/20.500.12128/12700
DOI: 10.3390/polym11121961
ISSN: 2073-4360